AirCurve 10 ST
Item # 37306
Technical Specifications
- Dimensions BiPAP (H x W x D): 4.6” x 6.5” x 5.7” @ 2.9 lbs
- Heated Humidifier (H4i) (H x W x D): 5.0” x 6.7” x 8.3” @ 1.6 lbs
- Symbols which appear on the device: Attention, consult accompanying documents; Drip proof; Type CF equipment; Dangerous voltage; Class II equipment; Start/Stop; Manufacturer.
- Operating pressure range: 3 to 25 cm H2O
- Maximum single fault steady state pressure: 40 cm H2O
- Pressure measurement tolerance: ±0.5 cm H2O ± 4% of the measured reading
- Flow measurement tolerance: ±0.1 or 20% of reading, whichever is greater
- VAuto mode: 4 to 25 cm H2O (measured at the mask); Min EPAP: 4 cm H2O; Max IPAP: 25 cm H2O; Pressure Support: 0 to 10 cm H2O S, ST and T modes IPAP: 4 to 25 cm H2O (measured at the mask); EPAP: 3 to 25 cm H2O (measured at the mask);
- Pressure Support: 0 to 22 cm H2O
- CPAP mode: 4 to 20 cm H2O (measured at the mask)
- DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871
- Sound pressure level: 26 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1: 2002
- 28 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007
- Sound power level: 36 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007
- Dimensions (L x W x H): 4.4” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)
- Weight: 2.9 lbs (1.3 kg)
- Power supply Input range :100–240 V, 50–60 Hz, 40 VA (typical power consumption), < 100 VA (maximum power consumption)
- Housing construction: Flame retardant engineering thermoplastic
- Operating temperature: 41ºF to 95ºF ( 5ºC to 35ºC)
- Operating humidity: 10–95% non-condensing
- Storage and transport temperature: -4ºF to 140ºF (-20ºC to 60ºC)
- Storage and transport humidity: 10–95% non-condensing
- Operating altitude: Sea level to 8,500’ (2,600 m)
- Supplemental oxygen: Recommended maximum supplemental oxygen flow: 4 L/min (VAuto mode); 15 L/min (CPAP, S, ST and T modes)
- Electromagnetic compatibility: Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments
- Air filter: Two-layered, powder-bonded, polyester non-woven fiber
- Air tubing: Flexible plastic, 1 x 6’6” (2 m)
- Air outlet: The 22 mm conical air outlet complies with ISO 5356-1
- IEC 60601-1 classification Class II (double insulation), Type CF
Note: The manufacturer reserves the right to change these specifications without notice.